search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

In vivo dosimetry during conformal radiotherapy: requirements for and findings of a routine procedure.

J H Lanson ,
M Essers ,
G J Meijer ,
A W Minken ,
G J Uiterwaal ,
B J Mijnheer

Abstract

MATERIALS AND METHODS

Based on pilot studies, routine in vivo dosimetry quality control (QC) protocols were implemented in the clinic. Entrance and exit diode dose measurements have been performed during two treatment sessions for 378 patients having prostate, bladder and parotid gland tumours. Dose calculations were performed with a CT-based three-dimensional treatment planning system. In our QC-protocol we applied action levels of 2.5% for the prostate and bladder tumour group and 4.0% for the parotid gland patients. When the difference between the measured dose at the dose specification point and the prescribed dose exceeded the action level the deviation was investigated and the number of monitor units (MUs) adjusted. Since an accurate dose measurement was necessary, some properties of the on-line high-precision diode measurement system and the long-term change in sensitivity of the diodes were investigated in detail.

CONCLUSIONS

Accurate in vivo dosimetry, using a diode measurement system, is a powerful tool to trace dosimetric errors during conformal radiotherapy in the range of 2.5-10%, provided that the system is carefully calibrated. The implementation of an intensive in vivo dosimetry programme requires additional staff for measurements and evaluation. The patient measurements add only a few minutes to the total treatment time per patient and guarantee an accurate dose delivery, which is a prerequisite for conformal radiotherapy.

RESULTS

The sensitivity of all diodes decreased by approximately 7% after receiving an integrated dose of 10 kGy, for 4 and 8 MV beams. For 34 (9%) patients the difference between the measured and calculated dose was larger than the action level. Systematic errors in the use of a new software release of the monitor unit calculation program, limitations of the dose calculation algorithms, errors in the planning procedure and instability in the performance of the accelerator have been detected.

PURPOSE

Conformal radiotherapy requires accurate knowledge of the actual dose delivered to a patient. The impact of routine in vivo dosimetry, including its special requirements, clinical findings and resources, has been analysed for three conformal treatment techniques to evaluate its usefulness in daily clinical practice.

More about this publication

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

Volume 52
Issue nr. 1
Pages 51-9
Publication date 01-07-1999

Full text links

Publisher website (DOI) 10.1016/s0167-8140(99)00074-2
Europe PubMed Central 10577686
Pubmed 10577686

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.