Abstract
METHODS
In this prospective feasibility study, 20 patients scheduled for RAPN were included from June 2024 to November 2025. During RAPN, tracked intraoperative ultrasound was used to align a patient-specific three-dimensional model with the kidney. This enabled real-time navigation of tracked surgical instruments relative to planned resection margins and automatic augmented reality visualization. We assessed the ability to achieve preoperatively planned resection volumes using IGN by computing the planned and resected tumor volume difference. Secondary outcomes were the navigation accuracy at the tumor, study measurement time, and usability of the IGN setup using the system usability scale (SUS).
CONCLUSION
IGN during RAPN is technically feasible and may enhance intraoperative tumor margin identification for precise resections, but wireless sensors and clinical evaluation in more complex, completely endophytic renal tumors are required in future.
RESULTS
In 70% (14/20) of the patients, the planned and resected volume difference was below 35%, indicating technical feasibility of the IGN technique. The median navigation accuracy was 2.2 (IQR 1.9-5.4) mm, the mean ± SD measurement time was 11 ± 3 min, and the median SUS score was 68 (IQR 65-70). Limitations of this study were wired sensors, one surgeon completed all SUS questionnaires, and limited clinical applicability in low-complex, exophytic tumors.
BACKGROUND
The aim of robot-assisted partial nephrectomy (RAPN) is complete tumor resection while maximally preserving healthy renal parenchyma. Image-guided navigation (IGN) provides real-time intraoperative guidance, which might be beneficial for endophytic margin assessment and resection of renal lesions. This study evaluates the technical feasibility and accuracy of IGN with electromagnetic tracking during RAPN.