search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Optimal combination of anti-scatter grids and software correction for CBCT imaging.

Uros Stankovic ,
Lennert S Ploeger ,
Marcel van Herk ,
Jan-Jakob Sonke

Abstract

METHODS

The scatter PSF was modeled as a sum of two bivariate Gaussian functions. The PSF parameters were estimated from a series of transmission measurements through polystyrene slabs of varying thickness with lead partial beam-blocker for three different ASG designs ranging from low (5.6), medium (9), and high (11) selectivity. The scatter correction scheme is based on iterative convolution of the current estimate of the primary with the scatter PSF until the root mean square deviation (RMSD) of the measured projection and the sum of the estimate of primary and scatter falls below a predefined threshold. The image quality was evaluated with the CIRS CBCT Image Quality and Electron Density phantom in a head and neck and pelvis configuration and the CIRS Virtual Male Human Patient. The image quality was quantified by the contrast-to-noise ratio (CNR) relative to the uncorrected scans and the root mean square deviation of the average gray values for different regions with respect to the nominal Hounsfield units and the mean difference of the reconstructed HU between the planning CT and CBCTs of the virtual human phantom.

CONCLUSIONS

The best scatter mitigation strategy was found to be a combination of a grid with selectivity larger than 9, combined with iterative scatter estimation. None of the investigated grids required increasing the imaging dose. The PSF determined using proposed method leads to image quality improvements results for all but one of the investigated scenarios.

RESULTS

For the head and neck phantom, the CNR increased with more advanced scatter correction algorithm and the ASG selectivity, reaching 3.9, 3.7, 3.5, and 3.1 for the high, medium, light, and with no grid configuration, respectively, combined with the iterative software correction. The same is true for the pelvis phantom with CNR improvement reaching 1.5 for the heavy and medium grid, 1.3 for the light grid, and 1.1 on its own. The HU RMSD for the head and neck phantom was 22 HU, 13 HU, 12 HU, and 6 HU for iterative correction without the grid, with the light grid, medium grid and the heavy grid, respectively. For same correction strategies, the values for the pelvis phantom where 170, 120, 34, and 27 HU. The average difference with the PCT of the virtual human phantom was 59 ± 48 HU and 63 ± 59 HU with scans reconstructed with the iterative correction and two higher selectivity grids. Visual inspection revealed similar trends for a head-and-neck and prostate cancer patient.

PURPOSE

Cone beam computed tomography (CBCT) has been widely adopted in clinical practice for image-guided radiotherapy. Soft tissue contrast and Hounsfield units are impaired to the presence of scattered radiation. In our previous work, we proposed a high selectivity anti-scatter grid (ASG) as a possible solution to the problem. An alternative approach is the application of iterative scatter correction using deconvolution with scatter point spread function (PSF). The purpose of this work was to compare the performance of ASGs with different selectivity with and without the iterative and uniform scatter corrections in terms of CBCT image quality. A secondary objective of this study was to develop a novel measurement approach to measure the scatter point spread functions.

More about this publication

Medical physics

Volume 44
Issue nr. 9
Pages 4437-4451
Publication date 01-09-2017

Full text links

Publisher website (DOI) 10.1002/mp.12385
Europe PubMed Central 28556204
Pubmed 28556204

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.