search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Simultaneous analysis of E1 and E2 by LC-MS/MS in healthy volunteers: estimation of reference intervals and comparison with a conventional E2 immunoassay.

Lennart J van Winden ,
Maik Kok ,
Manon Acda ,
Vincent Dezentje ,
Sabine Linn ,
Run-Zhang Shi ,
Huub H van Rossum

Abstract

Monitoring estrogen levels, especially estradiol (E2), is amongst others important for determining menopausal status and guidance of breast cancer treatment. We validated a serum E2 and estrone (E1) liquid chromatography tandem-mass spectrometry assay (LC-MS/MS) suitable for quantitation in human subjects. In addition, we compared our method with an E2 immunoassay (IA) and established preliminary reference values. Validation parameters were within the predetermined acceptance criteria. Assay linearity ranges were 4-1500 pmol/L for E1 and 4-2500 pmol/L for E2. Imprecision ranged from 7.4 to 9.6%. The lower limit of quantitation for E2 (8.0 pmol/L) was 11.4 times lower than the IA. The method comparison revealed differences in E2 quantitation up to 155% between both methods. The method allowed quantitation of E1 in all healthy volunteers, while E2 could not be detected in 95% versus 40% of the post-menopausal women using IA and LC-MS/MS, respectively. Male, pre-, peri- and postmenopausal female reference values were estimated. An LC-MS/MS based method combining E1 and E2 analysis was validated with superior E2 analytical sensitivity when compared to the IA.

More about this publication

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences

Volume 1178
Pages 122563
Publication date 15-07-2021

Full text links

Publisher website (DOI) 10.1016/j.jchromb.2021.122563
Europe PubMed Central 34224962
Pubmed 34224962

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.