search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Anatomy changes in radiotherapy detected using portal imaging.

Leah N McDermott ,
Markus Wendling ,
Jan-Jakob Sonke ,
Marcel van Herk ,
Ben J Mijnheer

Abstract

PATIENTS AND METHODS

Localisation images were obtained weekly prior to radiotherapy with an electronic portal imaging device (EPID). A series of 'difference images' was created by subtracting the first localisation image from that of subsequent fractions. Images from 81 lung, 40 head and neck and 34 prostate cancer patients were classified according to the changes observed. Changes were considered relevant if the average pixel value over an area of at least 1cm(2) differed by more than 5%, to allow for variations in linac output and EPID signal. Two patients were selected to illustrate the dosimetric effects of relevant changes. Their plans were re-calculated with repeat CT scans acquired after 4 weeks of treatment and compared with the difference images of the corresponding days.

CONCLUSIONS

Difference images are an efficient means of qualitatively detecting anatomical changes for various treatment sites in clinical practice. They can be used to identify changes for a particular patient, to indicate if the dose delivered to the patient would differ from planning and to detect if there is a need for re-planning.

RESULTS

Progressive changes were detected for 57% of lung and 37% of head and neck cancer patients studied. Random changes were observed in 37% of lung, 28% of head and neck and 82% of prostate cancer patients. For a lung case, an increase of 10.0% in EPID dose due to tumour shrinkage corresponded to an increase of 9.8% in mean lung dose. Gas pockets in the rectum region of the prostate case increased the EPID dose by 6.3%, and resulted in a decrease of the minimum dose to the planning target volume of 26.4%.

BACKGROUND AND PURPOSE

Localisation images normally acquired to verify patient positioning also contain information about the patient's internal anatomy. The aim of this study was to investigate the anatomical changes observed in localisation images and examples of dosimetric consequences.

More about this publication

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

Volume 79
Issue nr. 2
Pages 211-7
Publication date 01-05-2006

Full text links

Publisher website (DOI) 10.1016/j.radonc.2006.04.003
Europe PubMed Central 16698097
Pubmed 16698097

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.