search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiation.

Marcel van Herk ,
Lennert Ploeger ,
Jan-Jakob Sonke

Abstract

MATERIAL AND METHODS

The correction method is based on an alternating pulse sequence such that the kV source is enabled in only every other frame. MV scatter, noise and artifacts in the kV-unexposed frames are separated based on their spatial and temporal frequency characteristics and subsequently subtracted from the exposed frames. The method was tested on a phantom and one lung cancer patient scanned on an Elekta Synergy system simultaneous with a 9 Gy VMAT delivery.

CONCLUSIONS

A novel, efficient and effective method was developed to correct for MV scatter in CBCT scans acquired concurrent with rotational radiotherapy.

RESULTS

Without correction, acquisition during VMAT leads to significant cupping and loss of bone contrast. The novel method removes the signal due to MV scatter. Compared to acquisition without MV beam, the reconstruction quality is identical except for some unavoidable noise due to the scattered MV radiation in the exposed frames. Correction is performed in-line with acquisition without introducing delays.

BACKGROUND AND PURPOSE

Acquisition of cone-beam CT (CBCT) concurrent with VMAT results in scatter of the megavoltage (MV) beam onto the kilovoltage (kV) detector deteriorating CBCT image quality. The aim of this paper is to develop a method to estimate and correct for MV scatter reaching the kV panel.

More about this publication

Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

Volume 100
Issue nr. 3
Pages 365-9
Publication date 01-09-2011

Full text links

Publisher website (DOI) 10.1016/j.radonc.2011.08.019
Europe PubMed Central 21924785
Pubmed 21924785

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.