search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Accuracy assessment of wireless transponder tracking in the operating room environment.

Roeland Eppenga ,
Koert Kuhlmann ,
Theo Ruers ,
Jasper Nijkamp

Abstract

METHODS

Using a custom-made measurement setup, we assessed single transponder relative accuracy, absolute accuracy and jitter throughout the extended FoV at 1440 locations interspaced with 2.5 cm in each orthogonal direction. The NDI Polaris Spectra optical tracking system (OTS) was used as a reference. Measurements were taken in a room without surrounding distorting factors and repeated in an operating room (OR). In the OR, the influence of a carbon fiber and regular stainless steel OR tabletop was investigated.

CONCLUSION

This study shows that transponder tracking is promising for accurate tumor tracking in the operating room. This applies when using the standard FoV, but also when using the extended FoV up to 25 cm above the TA, substantially increasing flexibility.

RESULTS

The calibration of the OTS and transponder system resulted in an average root-mean-square error (RMSE) vector of 0.03 cm. For both the standard and extended FoV, all accuracy measures were dependent on transponder to tracking array (TA) distances and the absolute accuracy was also dependent on TA to OR tabletop distances. This latter influence was reproducible, and after calibrating this, the residual error was below 0.1 cm RMSE within the entire standard FoV. Within the extended FoV, this residual RMSE did not exceed 0.1 cm for transponder to TA distances up to 25 cm.

PURPOSE

To evaluate the applicability of the Calypso® wireless transponder tracking system (Varian Medical Systems Inc., USA) for real-time tumor motion tracking during surgical procedures on tumors in non-rigid target areas. An accuracy assessment was performed for an extended electromagnetic field of view (FoV) of 27.5 × 27.5 × 22.5 cm (which included the standard FoV of 14 × 14 × 19 cm) in which 5DOF wireless Beacon® transponders can be tracked.

More about this publication

International journal of computer assisted radiology and surgery

Volume 13
Issue nr. 12
Pages 1937-1948
Publication date 01-12-2018

Full text links

Publisher website (DOI) 10.1007/s11548-018-1838-z
Europe PubMed Central 30099659
Pubmed 30099659

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.