search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Multi-institutional dosimetric and geometric commissioning of image-guided small animal irradiators.

P E Lindsay ,
P V Granton ,
A Gasparini ,
S Jelveh ,
R Clarkson ,
S van Hoof ,
J Hermans ,
J Kaas ,
F Wittkamper ,
J-J Sonke ,
F Verhaegen ,
D A Jaffray

Abstract

METHODS

The system consists of a 225 kVp x-ray tube with fixed field size collimators ranging from 1 to 44 mm equivalent diameter. The x-ray tube is mounted opposite a flat-panel imaging detector, on a C-arm gantry with 360° coplanar rotation. Each institution performed a full commissioning of their system, including half-value layer, absolute dosimetry, relative dosimetry (profiles, percent depth dose, and relative output factors), and characterization of the system geometry and mechanical flex of the x-ray tube and detector. Dosimetric measurements were made using Farmer-type ionization chambers, small volume air and liquid ionization chambers, and radiochromic film. The results between the three institutions were compared.

CONCLUSIONS

With the exception of the half-value layer, and mechanical properties, there were significant differences between the dosimetric and geometric properties of the three systems. This underscores the need for careful commissioning of each individual system for use in radiobiological experiments.

RESULTS

At 225 kVp, with 0.3 mm Cu added filtration, the first half value layer ranged from 0.9 to 1.0 mm Cu. The dose-rate in-air for a 40 × 40 mm(2) field size, at a source-to-axis distance of 30 cm, ranged from 3.5 to 3.9 Gy/min between the three institutions. For field sizes between 2.5 mm diameter and 40 × 40 mm(2), the differences between percent depth dose curves up to depths of 3.5 cm were between 1% and 4% on average, with the maximum difference being 7%. The profiles agreed very well for fields >5 mm diameter. The relative output factors differed by up to 6% for fields larger than 10 mm diameter, but differed by up to 49% for fields ≤5 mm diameter. The mechanical characteristics of the system (source-to-axis and source-to-detector distances) were consistent between all three institutions. There were substantial differences in the flex of each system.

PURPOSE

To compare the dosimetric and geometric properties of a commercial x-ray based image-guided small animal irradiation system, installed at three institutions and to establish a complete and broadly accessible commissioning procedure.

More about this publication

Medical physics

Volume 41
Issue nr. 3
Pages 031714
Publication date 01-03-2014

Full text links

Publisher website (DOI) 10.1118/1.4866215
Europe PubMed Central 24593718
Pubmed 24593718

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.