search

menu

  • Research Research
    • Where science meets inspired minds

    • Back
    • Research
    • Our Science
    • Research Groups
    • Facilities & Platforms
    • Clinical research
    • Find a researcher
    • Publications
    • Knowledge Transfer
  • Careers & study Careers & study
    • Become a leader in cancer research

    • Back
    • Careers & study
    • Vacancies
    • Faculty
    • Scientific staff
    • Scientific support staff
    • Postdoctoral fellows
    • PhD Students
    • Operational staff
    • Clinical fellows
    • Life in Amsterdam
    • Student internships
  • News & Events News & Events
    • Check out our stories and events

    • Back
    • News & Events
    • News
    • Media & Press
    • Calendar
  • About us About us
    • Maximum impact for cancer patients

    • Back
    • About us
    • Our vision
    • Organization
    • Collaborations
    • Responsible Research
    • Support us
    • Visit us
    • Contact us
  • Support us
Support us
  • Home
  • Publications
  • Research
  • Publications
  • Article

Accurate in vivo dosimetry of a randomized trial of prostate cancer irradiation.

G J Meijer ,
A W Minken ,
K M van Ingen ,
B Smulders ,
H Uiterwaal ,
B J Mijnheer

Abstract

METHODS AND MATERIALS

Prostate patients are generally treated in our clinic with a 3-field isocentric technique: an 8-MV anteroposterior beam and 2 18-MV wedged laterals. All fields are shaped conformally to the PTV. Patients were randomized between two dose levels of 68 Gy and 78 Gy. During treatment, the entrance and exit dose were measured for each patient with diodes. Special 2.5-mm thick steel build-up caps were applied to make the diodes appropriate for measurements in 18-MV photon beams as well. Portal images were used to verify the correct position of the diodes and to detect and correct for gas filling in the rectum that may influence the exit dose reading. Entrance and exit dose measurements were converted to midplane dose, which was used in combination with a depth dose correction to obtain the dose at the specification point. An action level of 2.5% was applied.

CONCLUSION

Diodes can be used for accurate in vivo dosimetry during prostate irradiation in high-energy photon beams. The dose delivery in this randomized trial is guaranteed within the 2.5% limits on an individual patient basis. This could not be achieved without the in vivo dosimetry program, despite our high-standard quality assurance program of treatment delivery.

RESULTS

The added build-up for the diodes in the 18-MV beams resulted in correction factors that were only slightly sensitive to changes in beam setup and comparable to the corrections used in the 8-MV beams for diodes without extra build-up. The calibration factor increased almost linearly with cumulative dose: 0.7%/kGy for the 8-MV and 1.2%/kGy for the 18-MV photon beams. The introduction of average correction factors made the analysis easier, while keeping the accuracy within acceptable limits. In a period of 3 years, 225 patients were analyzed, from which 8 patients needed to be corrected. The average ratio of measured and prescribed dose was 1.009 (standard deviation [SD] 0.012) for the total group treated on two linear accelerators. When the results were analyzed per accelerator, the ratios were 1.002 (SD, 0.001) for Accelerator A and 1.015 (SD, 0.001) for Accelerator B. This difference could be attributed to the cumulative effect of three small imperfections in the performance of Accelerator B that were well within the limits of our quality assurance program.

PURPOSE

To guarantee an accurate dose delivery, within +/- 2.5%, in a Phase III randomized trial of prostate cancer irradiation (68 vs. 78 Gy) by means of a comprehensive in vivo dosimetry program.

More about this publication

International journal of radiation oncology, biology, physics

Volume 49
Issue nr. 5
Pages 1409-18
Publication date 01-04-2001

Full text links

Publisher website (DOI) 10.1016/s0360-3016(00)01556-x
Europe PubMed Central 11286849
Pubmed 11286849

Where science meets inspired minds

Contact

Plesmanlaan 121
1066CX Amsterdam

020 512 9111 communicatie@nki.nl

Quick links

  • Vacancies
  • News
  • Contact us
  • Media & Press

Follow us on

Disclaimer
Privacy statement
Cookies
Change cookie settings

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.