Development and validation of an UPLC-MS/MS method for the therapeutic drug monitoring of oral anti-hormonal drugs in oncology.

Abstract

A liquid chromatography-mass spectrometry assay was developed and validated for simultaneous quantification of anti-hormonal compounds abiraterone, anastrozole, bicalutamide, Δ(4)-abiraterone (D4A), N-desmethyl enzalutamide, enzalutamide, Z-endoxifen, exemestane and letrozole for the purpose of therapeutic drug monitoring (TDM). Plasma samples were prepared with protein precipitation. Analyses were performed with a triple quadrupole mass spectrometer operating in the positive and negative ion-mode. The validated assay ranges from 2 to 200 ng/mL for abiraterone, 0.2-20 ng/mL for D4A, 10-200 ng/mL for anastrozole and letrozole, 1-20 ng/mL for Z-endoxifen, 1.88-37.5 ng/mL for exemestane and 1500-30,000 ng/mL for enzalutamide, N-desmethyl enzalutamide and bicalutamide. Due to low sensitivity for exemestane, the final extract of exemestane patient samples should be concentrated prior to injection and a larger sample volume should be prepared for exemestane patient samples and QC samples to obtain adequate sensitivity. Furthermore, we observed a batch-dependent stability for abiraterone in plasma at room temperature and therefore samples should be shipped on ice. This newly validated method has been successfully applied for routine TDM of anti-hormonal drugs in cancer patients.

More about this publication

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
  • Volume 1106-1107
  • Pages 26-34
  • Publication date 01-02-2019